Inter-laboratory evaluation of instrument platforms and experimental workflows for quantitative accuracy and reproducibility assessment

dc.contributor.authorPercy, Andrew J.
dc.contributor.authorEspadas, Guadalupe
dc.contributor.authorSabidó Aguadé, Eduard, 1981-
dc.contributor.authorBorchers, Christoph H.
dc.date.accessioned2024-10-22T06:27:31Z
dc.date.available2024-10-22T06:27:31Z
dc.date.issued2015
dc.description.abstractThe reproducibility of plasma protein quantitation between laboratories and between instrument types was examined in a large-scale international study involving 16 laboratories and 19 LC–MS/MS platforms, using two kits designed to evaluate instrument performance and one kit designed to evaluate the entire bottom-up workflow. There was little effect of instrument type on the quality of the results, demonstrating the robustness of LC/MRM-MS with isotopically labeled standards. Technician skill was a factor, as errors in sample preparation and sub-optimal LC–MS performance were evident. This highlights the importance of proper training and routine quality control before quantitation is done on patient samples.
dc.format.mimetypeapplication/pdf
dc.identifier.citationPercy AJ, Tamura-Wells J, Albar JP, Aloria K, Amirkhani A, Araujo GD, et al. Inter-laboratory evaluation of instrument platforms and experimental workflows for quantitative accuracy and reproducibility assessment. EuPA Open Proteom. 2015 Sep;8:6-15. DOI: 10.1016/j.euprot.2015.06.001
dc.identifier.doihttp://dx.doi.org/10.1016/j.euprot.2015.06.001
dc.identifier.issn2212-9685
dc.identifier.urihttp://hdl.handle.net/10230/68293
dc.language.isoeng
dc.publisherElsevier
dc.relation.ispartofEuPA Open Proteomics. 2015 Sep;8:6-15
dc.rights© 2015 The Authors. Published by Elsevier GmbH. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subject.keywordAccuracy
dc.subject.keywordHuman plasma
dc.subject.keywordQuality control
dc.subject.keywordQuantitative proteomics
dc.subject.keywordReproducibility
dc.subject.keywordWorkflow validation
dc.titleInter-laboratory evaluation of instrument platforms and experimental workflows for quantitative accuracy and reproducibility assessment